Hostname: page-component-74656f9884-7drmj Total loading time: 0 Render date: 2026-11-05T05:44:13.180Z Has data issue: false hasContentIssue false

Experimental and numerical study of ultra-short laser-produced collimated Cu K α X-ray source

Published online by Cambridge University Press:  03 July 2017

R. Rathore*
Affiliation:
Laser Plasma Section, Raja Ramanna Centre for Advanced Technology, Indore 452013, Madhya Pradesh, India
V. Arora
Affiliation:
Laser Plasma Section, Raja Ramanna Centre for Advanced Technology, Indore 452013, Madhya Pradesh, India
H. Singhal
Affiliation:
Laser Plasma Section, Raja Ramanna Centre for Advanced Technology, Indore 452013, Madhya Pradesh, India
T. Mandal
Affiliation:
Laser Plasma Section, Raja Ramanna Centre for Advanced Technology, Indore 452013, Madhya Pradesh, India Homi Bhabha National Institute, Anushaktinagar, Mumbai 400 094, Maharashtra, India
J.A. Chakera
Affiliation:
Laser Plasma Section, Raja Ramanna Centre for Advanced Technology, Indore 452013, Madhya Pradesh, India Homi Bhabha National Institute, Anushaktinagar, Mumbai 400 094, Maharashtra, India
P.A. Naik
Affiliation:
Laser Plasma Section, Raja Ramanna Centre for Advanced Technology, Indore 452013, Madhya Pradesh, India Homi Bhabha National Institute, Anushaktinagar, Mumbai 400 094, Maharashtra, India
*
*Address correspondence and reprint requests to: R. Rathore, Laser Plasma Section, Raja Ramanna Centre for Advanced Technology, Indore 452013 MP, India. E-mail: ranjana@rrcat.gov.in

Abstract

K α X-ray sources generated from the interaction of ultra-short laser pulses with solids are compact and low-cost source of ultra-short quasi-monochromatic X-rays compared with synchrotron radiation source. Development of collimated ultra-short K α X-ray source by the interaction of 45 fs Ti:sapphire laser pulse with Cu wire target is presented in this paper. A study of the K α source with laser parameters such as energy and pulse duration was carried out. The observed K α X-ray photon flux was ~2.7 × 108 photons/shot at the laser intensity of ~2.8 × 1017 W cm−2. A model was developed to analyze the observed results. The K α radiation was coupled to a polycapillary collimator to generate a collimated low divergence (0.8 mrad) X-ray beam. Such sources are useful for time-resolved X-ray diffraction and imaging studies.

Information

Type
Research Article
Copyright
Copyright © Cambridge University Press 2017 
Figure 0

Fig. 1. Schematic representation of experimental setup. The X-ray CCD was used as dispersionless spectrograph when polycapillary collimator was not installed.

Figure 1

Fig. 2. X-ray spectrum recorded with dispersionless spectrograph. The spectrum shows Cu Kα (8.05 keV) and Kβ (8.9 keV) lines overriding on a bremsstrahlung background. The Kα X-ray flux was estimated from the area under the curve. The dashed line indicates the bremsstrahlung background level.

Figure 2

Fig. 3. Variation of Kα X-ray yield (a) with laser energy and (b) with laser pulse duration. Line in the figures show fitted flux distributions from modeling.

Figure 3

Fig. 4. Simulated electron density profile, generated by the interaction of pre-pulse with the target, using 1D hydrodynamic code (Helios, 2017) at different incident laser energies. The inset shows the ASE pedestal profile recorded by photo-diode (MRD 500) at 45 mJ laser energy.

Figure 4

Fig. 5. Simulated hot electron distributions (a) with laser energy at fixed pulse duration of 45 fs and (b) with laser pulse duration at constant laser energy of 23 mJ (fluence ~6.8 × 103 J cm−2).

Figure 5

Fig. 6. The knife edge intensity profile of the X-ray source (circles) with error function fit (solid line) and Gauss fit of its first derivative (dashed line). It shows a source size of ~155 µm, recorded at laser intensity of 2.2 × 1017 W cm−2.

Figure 6

Fig. 7. Variation of X-ray source size (a) with laser energy at fixed pulse duration of 45 fs and (b) with laser pulse duration at fixed laser energy of 36 mJ (fluence ~1.1 × 104 J cm−2). Source size shows monotonic increase with laser energy, whereas with increase in pulse duration the source size first decreases and then it increases slowly.

Figure 7

Fig. 8. Collimated X-ray beam profile.

Figure 8

Fig. 9. The diffraction patterns from HOPG crystal (a) using collimated X-ray beam. The inset shows the image of diffracted beam and (b) using diverging X-ray beam. The dashed line indicates the bremsstrahlung background. It is evident that the bremsstrahlung noise is decreased substantially in Fig. 9a.