To segment or not to segment the PTV in SIB-VMAT of head-and-neck cancer: a dosimetric comparison with relevance to clinical practice

Published online by Cambridge University Press:  09 March 2021

Manu Mathew
Affiliation:
Department of Radiation Oncology, Dr Ida B. Scudder Cancer Center, Christian Medical College, Vellore, Tamil Nadu, India
Henry F. Godson
Affiliation:
Department of Radiation Oncology, Dr Ida B. Scudder Cancer Center, Christian Medical College, Vellore, Tamil Nadu, India
Abel J. Thomas
Affiliation:
Department of Radiation Oncology, Dr Ida B. Scudder Cancer Center, Christian Medical College, Vellore, Tamil Nadu, India
Sasidharan Balu Krishna*
Affiliation:
Department of Radiation Oncology, Dr Ida B. Scudder Cancer Center, Christian Medical College, Vellore, Tamil Nadu, India
Sathish Kumar A
Affiliation:
Department of Radiation Oncology, Dr Ida B. Scudder Cancer Center, Christian Medical College, Vellore, Tamil Nadu, India
Rajesh Isiah
Affiliation:
Department of Radiation Oncology, Dr Ida B. Scudder Cancer Center, Christian Medical College, Vellore, Tamil Nadu, India
Subhashini John
Affiliation:
Department of Radiation Oncology, Dr Ida B. Scudder Cancer Center, Christian Medical College, Vellore, Tamil Nadu, India
Simon P. Pavamani
Affiliation:
Department of Radiation Oncology, Dr Ida B. Scudder Cancer Center, Christian Medical College, Vellore, Tamil Nadu, India
*
Author for correspondence: Dr Sasidharan Balu Krishna, Department of Radiation Oncology, Unit 2, Dr Ida B. Scudder Cancer Center, Christian Medical College, Vellore, Tamil Nadu 632004, India. Tel: 9626262296. E-mail: balunair@cmcvellore.ac.in
Rights & Permissions [Opens in a new window]

Abstract

Aim:

This study aimed to evaluate and compare simultaneous integrated boost-based volumetric modulated arc therapy (SIB-VMAT) of head-and-neck plans optimised using segmented and non-segmented intermediate-risk target volumes.

Materials and methods:

CT data of 20 patients with locally advanced laryngeal cancer treated with radical chemoradiation were included retrospectively. Both segmented [planning target volume (PTV) IR!] and non-segmented PTV (PTV IR) volumes were created for the intermediate-risk volume. Correspondingly, two VMAT plans were generated for every CT dataset. Dosimetry parameters obtained from cumulative dose volume histogram and the quality indices such as conformity and homogeneity indices were evaluated for both plans and were statistically analysed.

Results:

Maximum dose of PTV IR! was observed to be higher in the non-segmented plans (7281·45 versus 7075·75 cGy) and was statistically significant (p = 0·002). Homogeneity index (HI) of PTV IR! in segmented plans fared better compared to non-segmented plans (0·1 versus 0·12, p = 0·01). All other dosimetry parameters were found to be similar in both plans.

Conclusion:

This study shows that using segmented volumes for planning will lead to more homogenous plans with regard to intermediate- and low-risk volumes, especially under controlled settings.

Information

Type
Original Article
Copyright
© The Author(s), 2021. Published by Cambridge University Press
Figure 0

Figure 1. Showing the two types of segmentation for intermediate- and low-risk volumes used for optimisation. (a) Non-segmented volume (PTV IR = PTV IR + PTV HR) and (b) segmented volume (PTV IR! = PTV IR - PTV HR with a 2 mm gap).

Figure 1

Figure 2. A representative cumulative DVH showing the dosimetry of segmented (PTV_IR!) (red) and non-segmented (PTV_IR) (blue) volumes. The elephant foot noted in the non-segmented volumes is due to overlap with high-risk PTV.

Figure 2

Figure 3. Representative CT images of a case of laryngeal cancer. Both panels represent the same slice. (a) Segmented PTV IR! is shown in red and PTV HR in orange and there is a 2 mm gap between PTV structures. (b) Non-segmented PTV IR is in blue colour with a Boolean addition of PTV HR in orange.

Figure 3

Table 1. Analysis of dose volume parameters of segmented and non-segmented plans. Level of significance by paired t-test

Figure 4

Figure 4. Variation in dose spillage with the 95% of the prescription dose of PTV HR in non-segmented plans (a) and segmented plans (b). The dose colour wash set here for a minimum dose of 66·5 Gy and above.

Figure 5

Table 2. Analysis of the quality indices of both segmented and non-segmented plans

Figure 6

Table 3. Comparison of the segmented and non-segmented plans in SIB IMRT based on our study results