Hostname: page-component-74656f9884-c52kr Total loading time: 0 Render date: 2026-11-07T10:07:49.395Z Has data issue: false hasContentIssue false

Comparison of different excitation modes for the analysis of light elements with a TXRF vacuum chamber

Published online by Cambridge University Press:  22 May 2015

J. Prost*
Affiliation:
Atominstitut, Vienna University of Technology, Stadionallee 2, 1020 Vienna, Austria
P. Wobrauschek
Affiliation:
Atominstitut, Vienna University of Technology, Stadionallee 2, 1020 Vienna, Austria
C. Streli
Affiliation:
Atominstitut, Vienna University of Technology, Stadionallee 2, 1020 Vienna, Austria
*
a) Author to whom correspondence should be addressed. Electronic mail: josef.prost@ati.ac.at

Abstract

The aim of this work was to compare different excitation modes for the analysis of light elements from carbon (Z = 6) upwards using a total reflection X-ray fluorescence analysis (TXRF) vacuum chamber which allows the attachment of different X-ray tubes and detectors. In the first set of experiments, two water-cooled high-power X-ray tubes with Cr (Z = 25) and Cu (Z = 29) anodes, respectively, were compared with an air-cooled low-power tube with Mo anode (Z = 42) and a thin Be window for the transmission of Mo-L lines. In the first two cases, monochromatic Kα radiation was used for excitation, while in the case of the Mo tube the multilayer acted as a cut-off reflector and part of the Mo bremsstrahlung continuum together with the Mo-L series were used for excitation. Multi-element standards containing elements ranging from Na (Z = 11) to Ti (Z = 22) were analyzed by a silicon drift detector (SDD) with a 300 nm ultrathin polymer window (UTW). Detection limits were calculated and compared for the three excitation modes. The second set of experiments was performed using an air-cooled low-power X-ray tube with Rh anode (Z = 45) in order to show that a conventional SDD with a 25 μm beryllium window can be used for the detection of elements from Na upwards. The use of compact air-cooled low-power X-ray tubes together with Peltier-cooled SDDs with UTW should lead to the development of highly sensitive tabletop vacuum TXRF spectrometers with a design optimized for the analysis of light elements. Detection limits as achieved by vacuum chambers using conventional water-cooled high-power tubes (e.g. Streli et al., 2004) are realistically achievable with the new approach.

Information

Type
Technical Articles
Copyright
Copyright © International Centre for Diffraction Data 2015 
Figure 0

Figure 1. (Color online) Atominstitut TXRF Standard Vacuum Chamber. Panel (a) shows a top view of the chamber. On the left side different X-ray sources can be attached to the chamber, either via the tube housing shown in the picture or via special adapter pieces. The detector feed-through is shown on top of the photograph. Monochromator unit and sample stage can be seen inside the vacuum chamber. Panel (b) shows a close-up view of the detector feed-through, and on (c) the Mo anode low-power tube connected to the chamber via an adapter piece can be seen.

Figure 1

Table I. Source–detector combinations applied in the experiments of this study. Two water-cooled high-power diffraction tubes were compared with two air-cooled low-power tubes. Furthermore, two different silicon drift detectors were used, one with an ultrathin polymer window but only 7 mm2 active area, the other one with a 25 μm beryllium window, and 80 mm2 active area.

Figure 2

Figure 2. (Color online) Detection limits of all measured samples. Na, Mg, and Al could be evaluated for all three concentrations (50, 100, and 200 ng), while S, K, Ca, and Ti were present only in the 50 and 100 ng samples. With Mo cut-off excitation, Na was below the limit of detection for the 50 ng sample.

Figure 3

Figure 3. (Color online) Overlay of spectra from the 50 ng multielement sample. Cr excitation leads to the highest intensities over an excellent background. Cu also provides good background conditions, but much weaker peak intensities. The intensities obtained in the Mo cut-off excitation mode are about an order of magnitude lower, but the detection of Mg and Al was possible, because the spectrum is nearly background-free.

Figure 4

Figure 4. (Color online) Overlay of spectra from the 200 ng sample containing only Na, Mg, and Al. It can be seen that the peaks for the three elements are clearly resolved. In every spectrum, oxygen is visible. With Cr excitation a small carbon peak is also visible.

Figure 5

Table II. Detection limits (DL) (1000 s) in ng calculated for the 50 ng multielement sample, obtained with a 7 mm2 silicon drift detector with ultrathin window, comparing different excitation sources (900 W Cr, 800 W Cu, and 50 W Mo cut-off). It can be seen that the excitation with Cr leads to the best results. Mg and Al could be detected with all the three excitation sources, whereas Na was below the detection limit for the Mo cut-off mode.

Figure 6

Figure 5. (Color online) Spectrum of the 100 ng Na, Mg, Al, S, and Ti multielement sample. Peaks of Na, Mg, and Al are clearly visible, even though a detector with 25 μm beryllium window was used.

Figure 7

Table III. Detection limits (DL) (1000 s) in ng obtained from the 100 ng multielement sample (Na, Mg, Al, S, K, Ca, and Ti) with Rh excitation and an 80 mm2 silicon drift detector with a 25 μm beryllium entrance window.