High-energy X-ray applications: current status and new opportunities

Published online by Cambridge University Press:  30 January 2018

Dubravka Šišak Jung*
Affiliation:
DECTRIS Ltd, Täfernweg 1, 5405 Baden-Dättwil, Switzerland
Tilman Donath
Affiliation:
DECTRIS Ltd, Täfernweg 1, 5405 Baden-Dättwil, Switzerland
Oxana Magdysyuk
Affiliation:
Diamond Light Source, Didcot, United Kingdom
Jozef Bednarcik
Affiliation:
Deutsches Elektronen-Synchrotron, Notkestrasse 85, Hamburg, Germany
*
a) Author to whom correspondence should be addressed. Electronic mail: dubravka.sisak@dectris.com

Abstract

Characterization of semi and noncrystalline materials, monitoring structural phase transitions in situ, and obtaining structural information together with spatial distribution of the investigated material are only a few applications that hugely benefitted from the combination of high-energy X-rays and modern algorithms for data processing. This work examines the possibility of advancing these applications by shortening the data acquisition and improving the data quality by using the new high-energy PILATUS3 CdTe detector.

Information

Type
Technical Articles
Copyright
Copyright © International Centre for Diffraction Data 2018 
Figure 0

Figure 1. (Colour online) Individual optimization of two technologies (sensor, readout) is an underlying principle of HPC design and a key to specific detector properties.

Figure 1

Figure 2. (Colour online) (a) In comparison with Si (pink dashed), CdTe sensor (blue line) shows higher absorption efficiency at high energies. (b) Quantum efficiency of PILATUS3 CdTe, as measured (dots) and simulated (blue dashed line).

Figure 2

Figure 3. (Colour online) Count rate linearity of the PILATUS3 CdTe reaches 5 × 106 counts/s/pixel. Measured values (circles) fit to the simulated values (line). As a comparison, a response of ideal detector (dashed line) is shown.

Figure 3

Figure 4. (Colour online) PXRD pattern (t = 1 ms) reveals identifiable low-intensity peaks and average background close to zero (ID12, DLS).

Figure 4

Figure 5. (Colour online) An exposure time as low as 0.005 ms allows scattering signal to be detected (pink), while 1 ms exposure time results in high-quality scattering data (red), (HPCAT, APS).

Figure 5

Figure 6. (Colour online) Standard deviation of the residual curve drops exponentially with the number of merged patterns.

Figure 6

Figure 7. (Colour online) Single scattering pattern (t = 0.1 s, σ = 0.07) can be used for calculations of PDF up to 18 Å.