Cartilage ‘shoe’: a new technique for stabilisation of titanium total ossicular replacement prosthesis at centre of stapes footplate

Published online by Cambridge University Press:  08 March 2017

D Beutner
Affiliation:
Department of Otorhinolaryngology, Head and Neck Surgery, University of Cologne, Germany
J C Luers
Affiliation:
Department of Otorhinolaryngology, Head and Neck Surgery, University of Cologne, Germany
K B Huttenbrink*
Affiliation:
Department of Otorhinolaryngology, Head and Neck Surgery, University of Cologne, Germany
*
Address for correspondence: Prof Karl-Bernd Huttenbrink, Professor and Chairman, Department of Otorhinolaryngology, Head and Neck Surgery, University of Cologne, Kerpener Strasse 62, 50924 Cologne, Germany. Fax: +49 221 478 4793 E-mail: huttenbrink.k-b@uni-koeln.de
Rights & Permissions [Opens in a new window]

Abstract

Objectives:

After tympanoplasty using a total ossicular replacement prosthesis, many unsatisfactory hearing results are due to dislocation of the prosthesis.

Material and methods:

We developed a cartilage guide for stabilising the total ossicular replacement prosthesis in the oval window niche. An oval-shaped piece of cartilage measuring 2.5 × 3.5 mm with a central hole was precisely punched out of a thin cartilage plate. The cartilage was placed in the oval niche, and its hole centred the prosthesis on the stapes footplate.

Results:

Hearing results in 52 patients confirmed acoustically the effectiveness of this method of total ossicular replacement prosthesis stabilisation on the stapes footplate. Subsequent ‘second-look’ surgery revealed stable ingrowth of the cartilage ‘shoe’ into the oval niche.

Conclusion:

Such a cartilage shoe might address one of the causes of unsatisfactory hearing following ossicular chain reconstruction with a total ossicular replacement prosthesis.

Information

Type
Main Article
Copyright
Copyright © JLO (1984) Limited 2008
Figure 0

Fig. 1 (a) The cartilage punch used to produce the cartilage ‘shoe’, consisting of an outline punch, a centre hole punch, a spring and an ejection needle. (b) The cartilage shoe, incorporating the prosthesis foot of a titanium total ossicular replacement prosthesis. (c) The anterior and posterior poles of the cartilage shoe can be cut away in cases with persistent crura remnants.

Figure 1

Table I Patient and Surgical Data

Figure 2

Fig. 2 (a) Intra-operative view of the cartilage ‘shoe’ in the oval window niche, guiding a total ossicular replacement prosthesis (TORP) onto the footplate centre. (b) View at revision surgery, showing a reliable TORP position, without pathological findings.

Figure 3

Fig. 3 Patients’ pre- and post-operative audiometric results, for use of a cartilage ‘shoe’ in type III tympanoplasty (n = 52).

Figure 4

Table II Patients’ Audiometric Results*