The relevance of electrocardiogram filtering in the differential diagnosis of Brugada syndrome: An illustrative case report

Published online by Cambridge University Press:  01 December 2023

Carlos Baena-Palomino*
Affiliation:
Pediatric Cardiology Division, Puerta del Mar University Hospital, Cádiz, Spain
Ana Castellano-Martínez
Affiliation:
Pediatric Nephrology Division, Puerta del Mar University Hospital, Cádiz, Spain
Moisés Rodríguez-González
Affiliation:
Pediatric Cardiology Division, Puerta del Mar University Hospital, Cádiz, Spain
*
Corresponding author: C. Baena-Palomino; Email: cbaenap@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

We present a clinical case of a teenager in whom a Brugada syndrome electrical pattern was found which was finally diagnosed as a Brugada phenocopy due to a wrong setup of the high-pass filter on the electrocardiogram

Information

Type
Brief Report
Copyright
© The Author(s), 2023. Published by Cambridge University Press
Figure 0

Table 1. Brugada phenocopy categories and diagnostic criteria (adapted from Anselm et al.)5

Figure 1

Figure 1. a and b. 12-lead electrocardiogram performed manually at the centre of origin with an analog electrocardiogram with a high-pass filter which had a frequency of 0.67 hz (lower left area of Fig. 1a). Right precordial leads (V2-V3) show convex ST-segment elevations ≥2 mm with respect to the isoelectric line of base, with saddleback morphology. c. 12-lead electrocardiogram performed in our centre with digital electrocardiogram and high-pass filter with a frequency of 0.05 hz (upper right area of Fig. 1c), in which no abnormalities were observed. Note that the applied 45 hz low-pass filter in Figure 1c is also incorrect but did not influence the diagnosis of our patient, American Heart Association recommend a minimum of 150 hz for the low-pass filter. ECM: myogram or muscle filter. DFT0: drift filter. Note that DFT could refer to any filter, but the value 0,67 hz is usually a configuration used for high-pass filters. Hz = hertz.