Hostname: page-component-74656f9884-j7cmm Total loading time: 0 Render date: 2026-11-07T13:36:24.949Z Has data issue: false hasContentIssue false

Tunable glass reference materials for quantitative backscattered electron imaging of mineralized tissues

Published online by Cambridge University Press:  21 August 2012

Sara E. Campbell
Affiliation:
Department of Mechanical Engineering, University of Colorado, Boulder, Colorado 80309; and Materials Reliability Division, NIST, Boulder, Colorado80305
Roy H. Geiss
Affiliation:
Materials Reliability Division, NIST, Boulder, Colorado80305
Steve A. Feller
Affiliation:
Physics Department, Coe College, Cedar Rapids, Iowa52402
Virginia L. Ferguson*
Affiliation:
Department of Mechanical Engineering, University of Colorado, Boulder, Colorado80309
*
a)Address all correspondence to this author. e-mail: virginia.ferguson@colorado.edu

Abstract

Backscattered electron microscopy provides gray-level contrast resulting from variations in atomic composition. Through the use of reference materials, quantitative backscattered electron (qBSE) imaging can be used to measure the mineral content of mineralized tissues at submicron resolution. We have developed novel tunable reference materials that can be adjusted for analysis of an individual tissue or a wide range of tissues with variable atomic density. As an alternative to conventional metallic reference materials, these amorphous materials maintain long-term stability and possess no long-range order that may induce channeling contrast. Using these reference materials, we characterized the mineral content of a broad range of mineralized tissues from immature mouse femur to whale bulla. Mineral volume fraction correlated to more traditional measurements of mineral content with microcomputed tomography and ashing techniques. Further, we demonstrate the advantage of location-matched measurements of nanomechanical properties and qBSE mineral content.

Information

Type
Articles
Copyright
Copyright © Materials Research Society 2012
Figure 0

TABLE I. Zcomp and ηcomp for calibration materials and glasses calculated using Eqs. (4) and (5), respectively.

Figure 1

FIG. 1. Raw GL histograms of the five glass reference materials and those from cow bone and rabbit enamel. The BSE image (top, scale bar 200 μm) shows that the glasses generally increase in GL with increasing X values. Glass references with X = 0 and X = 0.15 were selected for this study because they encompass the pertinent range of GLs.

Figure 2

FIG. 2. WMGL of calibration materials and glass references versus (a) mean atomic number Z and (b) qBSE coefficient ηcomp. Calculations for Zcomp of composite materials with low atomic number elements do not correctly predict backscattered behavior, as seen in (a). However, calculations for ηcomp do correctly relate to backscattered behavior (i.e., WMGL) as seen in (b).

Figure 3

FIG. 3. (a) Relationship of ηcomp to calculated MV%. (b) Calibration curve for conversion of WMGL values to mineral volume percent, MV%.

Figure 4

FIG. 4. Calibrated qBSE images of representative samples displaying variation of mineral content both within and between samples: (a) 2-wk-old mouse femur, (b) young lamb femur, (c) carp dermal bone, (d) 20-wk-old mouse femur, (e) cow femur, and (f) whale bulla. Increasing mineral content from (a) to (f) is evident with a general increase in WMGL. All images were taken at 500x (scale bar = 100 μm), at 20 kV acceleration voltage, and a 15-mm working distance.

Figure 5

FIG. 5. Mineral volume percent of the mineralized tissues measured by qBSE correlated to (a) ash mineral volume percent and (b) microCT density.

Figure 6

FIG. 6. Location-matched measurements of nanoindentation tests and qBSE MV% on young lamb femur. (a) qBSE image showing an indentation array of 5 × 4 indents with 20 μm spacing between test sites. Test sites are categorized as interstitial bone (circles) or osteonal bone (diamonds). The test site with a X though the circle was removed from analysis because the indenter tip landed on an osteocyte lacuna. (b) Corresponding nanoindentation modulus for each test site versus MV% analyzed within the area surrounding each test site.

Figure 7

FIG. 7. The log of plane strain modulus E′ versus mineral volume fraction MV% of all mineralized tissue samples for each indentation site. The solid lines are Voigt–Reuss bounds, and the dashed lines are the Hashin–Shtrikman composite bounds.