Temperature dependent dielectric spectroscopy of muscle tissue phantom

Published online by Cambridge University Press:  19 March 2020

Ondrej Fiser*
Affiliation:
Department of Biomedical Technology, Faculty of Biomedical Engineering, CTU in Prague, Prague, Czech Republic
Sebastian Ley
Affiliation:
Faculty of Computer Science and Automation, Institute of Biomedical Engineering and Informatics, TU Ilmenau, Ilmenau, Germany
Marko Helbig
Affiliation:
Faculty of Computer Science and Automation, Institute of Biomedical Engineering and Informatics, TU Ilmenau, Ilmenau, Germany
Jürgen Sachs
Affiliation:
Ilmsens GmbH, Ilmenau, Germany
Michaela Kantova
Affiliation:
Department of Electromagnetic Field, Faculty of Electrical Engineering, CTU in Prague, Prague, Czech Republic
Jan Vrba
Affiliation:
Department of Electromagnetic Field, Faculty of Electrical Engineering, CTU in Prague, Prague, Czech Republic
*
Author for correspondence: Ondrej Fiser, E-mail: Ondrej.fiser@fbmi.cvut.cz
Rights & Permissions [Opens in a new window]

Abstract

The temperature dependence of the dielectric parameters of tissues and tissue-mimicking phantoms is very important for non-invasive temperature measurement in medical applications using microwaves. We performed measurements of this dependence in the temperature range of 25–50°C using distilled water as a reference liquid commonly used in dielectric property studies. The results were compared with the literature model in the frequency range of 150–3000 MHz. Using this method, the temperature dependence of dielectric parameters of a new muscle tissue-mimicking phantom based on agar, polyethylene powder, and polysaccharide material TX-151 was measured in the temperature range of 25–50°C. The temperature dependence of the dielectric properties of this new muscle phantom was fitted to that of the two-pole Cole–Cole model and the deviation of the results between measured and modeled data was quantified.

Information

Type
Research Paper
Copyright
Copyright © Cambridge University Press and the European Microwave Association 2020
Figure 0

Fig. 1. Measurement setup for the measurement of temperature dependence of tissue mimicking phantoms (a) and the photo of the water container with the coaxial probe and MUT (b).

Figure 1

Fig. 2. Measured temperature-dependent relative permittivity (a)* and effective conductivity (b) with combined standard uncertainty uC (equation (1)) of distilled water (solid lines) in the frequency range of 150–3000 MHz after temperature-dependent calibration compared to the model from Ellison marked as E – (dashed lines).

Figure 2

Fig. 3. Differences between the Ellison model and measured temperature-dependent relative permittivity (a) and effective conductivity (b) depending on the calibration procedure. Dashed lines corresponding with the measurement without temperature dependent calibration and solid lines with temperature calibration.

Figure 3

Table 1. The composition of proposed muscle phantom, substances in percentages by weight.

Figure 4

Fig. 4. Measured temperature dependency of the new muscle tissue phantom in the temperature range of 25–50°C.

Figure 5

Fig. 5. (a) Relative permittivity and (b) effective conductivity of tissue-mimicking phantom modeled from the fitted two-pole Cole–Cole model.

Figure 6

Table 2. Temperature coefficients of the two-pole Cole–Cole model of the muscle tissue phantom

Figure 7

Fig. 6. Differences between measured and modeled data by the two-pole Cole–Cole model of (a) relative permittivity and (b) effective conductivity.