Hostname: page-component-74656f9884-54qdv Total loading time: 0 Render date: 2026-11-10T05:23:40.511Z Has data issue: false hasContentIssue false

Verification of dose distribution by different material properties in intraoral mold irradiation

Published online by Cambridge University Press:  20 July 2023

Hiraku Fuse*
Affiliation:
Department of Radiological Sciences, Ibaraki Prefectural University of Health Sciences, Ibaraki, Japan
Fumihiro Tomita
Affiliation:
Graduate School of Health Sciences, Department of Radiological Sciences, Ibaraki Prefectural University of Health Sciences, Ibaraki, Japan
Kenji Yasue
Affiliation:
Department of Radiological Sciences, Ibaraki Prefectural University of Health Sciences, Ibaraki, Japan
Hideaki Ikoma
Affiliation:
Department of Radiation Technology, Ibaraki Prefectural Central Hospital, Ibaraki, Japan
Shin Miyakawa
Affiliation:
Department of Radiological Sciences, Ibaraki Prefectural University of Health Sciences, Ibaraki, Japan
Norikazu Kori
Affiliation:
Department of Radiological Sciences, Ibaraki Prefectural University of Health Sciences, Ibaraki, Japan
Tatsuya Fujisaki
Affiliation:
Department of Radiological Sciences, Ibaraki Prefectural University of Health Sciences, Ibaraki, Japan
Yoshiyuki Ishimori
Affiliation:
Department of Radiological Sciences, Ibaraki Prefectural University of Health Sciences, Ibaraki, Japan
Masahiko Monma
Affiliation:
Department of Radiological Sciences, Ibaraki Prefectural University of Health Sciences, Ibaraki, Japan
Toshiyuki Okumura
Affiliation:
Department of Radiation Technology, Ibaraki Prefectural Central Hospital, Ibaraki, Japan
Yoshio Tamaki
Affiliation:
Department of Radiation Oncology, Fukushima Rosai Hospital, Iwaki, Fukushima, Japan
*
Corresponding author: Hiraku Fuse; Email: fuseh@ipu.ac.jp
Rights & Permissions [Opens in a new window]

Abstract

Background:

Brachytherapy is an effective local treatment for early-stage head and neck cancers. Mold irradiation is a method in which the source is placed in the oral cavity in sites where the soft tissue is thin and an irradiation source cannot be implanted. However, dose calculations based on TG-43 may be subject to uncertainty due to the heterogeneity of tissues and materials used for the irradiation of head and neck cancers.

Materials and Methods:

In this study, we investigated the basic physical properties of different materials and densities in the molds, retrospectively analysed patient plans and verified the doses of intraoral mold irradiation using a dose verification system with MC simulations specifically designed for brachytherapy, which was constructed independently.

Results and Discussion:

Dose–volume histograms were obtained with a treatment planning system (TG-43) and MC simulation and revealed a non-negligible difference in coverage of high-risk clinical target volume (HR-CTV) and organ at risk (OAR) between calculations using computed tomography values and those with density changes. The underdose was 10·6%, 3·7% and 5·6% for HR-CTV, gross tumour volume and OAR, respectively, relative to the treatment plan. The calculations based on the differences in the elemental composition and density changes in TG-43, a water-based calculation algorithm, resulted in clinically significant dose differences. The validation method was used only for the cases of complex small source therapy.

Conclusion:

The findings of this study can be applied to more complex cases with steeper density gradients, such as mold irradiation.

Information

Type
Original Article
Copyright
© The Author(s), 2023. Published by Cambridge University Press
Figure 0

Table 1. Elemental composition, relative electron density and physical density of material used in molds

Figure 1

Figure 1. Geometry for calculating dose profiles using MC simulation. The source was placed in the center of a water sphere with a radius of 40 cm. The slab phantom was 1 cm thick. Dose profiles were obtained in the direction of the short axis of the source.

Figure 2

Figure 2. The images of mold applicator used for brachytherapy of soft palate cancer with A) cranial, B) right lateral, C) caudal, and D) posterior views. The clear resin is BS and the pink resin is OSTRON.

Figure 3

Figure 3. Three cross-sections (axial, sagittal and coronal) of the high-risk clinical target volume (HR-CTV; light blue) and organ-at-risk (OAR; purple) ROIs. The red arrows indicate the mold. High computed tomography values in the mold indicate catheters for transporting 192Ir sources.

Figure 4

Figure 4. The dose distribution (solid line) and dose difference (dash line) with TG-43 due to the varying density of the slab phantom. The upper figure shows the calculation results by MC simulation using the slab phantom of BS, and the lower figure shows the calculation results by MC simulation using the slab phantom of OSTRON.

Figure 5

Figure 5. Tissue-absorbed dose distribution calculated using MC SIMULATION and water-absorbed it using TPS (TG-43). The thin and thick lines indicate the MC SIMULATION and TG-43 results, respectively. MC simulations were performed based on CT values (upper left) and molds overwritten to 1·0 g/cm3 (upper right) and 1·5 g/cm3 (lower left).

Figure 6

Figure 6. Dose–volume histogram (DVH) obtained using the treatment planning system (TPS; TG-43) and three patterns for Monte Carlo simulation with constant CT values and calculation with minimum and maximum densities of 1·0 and 1·5 g/cm3 overwritten to reflect density changes due to fabrication.

Figure 7

Table 2. D98%, D90%, D2ccV100%, V150%, V200% and DHI calculated based on mold density obtained from CT value and results reflecting the mold density differences due to fabrication