Impact of software changes: Transit dose and source position accuracy of the Eckert & Ziegler BEBIG GmbH MultiSource® high dose rate (HDR) brachytherapy treatment unit

Published online by Cambridge University Press:  02 August 2012

A. Palmer*
Affiliation:
Radiotherapy Physics, Medical Physics Department, Portsmouth Hospitals NHS Trust, United Kingdom Department of Physics, Faculty of Engineering and Physical Science, University of Surrey, United Kingdom
*
Correspondence to: Antony Palmer, Head of Radiotherapy Physics, Medical Physics Department (Radiotherapy), F-Level, Queen Alexandra Hospital, Portsmouth Hospitals NHS Trust, Portsmouth, United Kingdom. E-mail: antony.palmer@porthosp.nhs.uk
Rights & Permissions [Opens in a new window]

Abstract

Purpose: Medical device performance checks are essential following changes to control system software. This work investigates the effects of new software on the performance of a high dose rate (HDR) brachytherapy treatment unit.

Methods and Materials: A performance assessment was undertaken of the Eckert & Ziegler BEBIG GmbH MultiSource® HDR treatment unit following software upgrade. Video recordings of source transits were used to calculate transit doses, and autoradiography used to measure source dwell positions. Results were compared to a previous study.1

Results: All results showed improved performance with the new compared to old control software. Optimal source movement profiles were observed with maximum transit speeds of 63 (+/−4) mm s−1 between dwells of 5.0 mm separation. The maximum error in transit dose correction with the new software was 2.5 % at 10.0 mm perpendicular from the source axis, compared to 5.6 % previously. The new software eliminated a causal relationship between curvature of the source transfer tubes and dwell position uncertainty.

Conclusions: This work demonstrates the need for comprehensive medical device system checks following software changes. Technical improvements in HDR device performance have been achieved with the new software; reducing transit doses, improving transit dose correction, and improving source positioning accuracy.

Information

Type
Technical Note
Copyright
Copyright © Cambridge University Press 2013
Figure 0

Figure 1. Transit dose calculation points, D10 and D20, at 10 mm and 20 mm respectively, perpendicular distance from the centre of the intended dwell position.

Figure 1

Figure 2. Position and speed of source during transit from the EZ BEBIG Multisource® on approach to the first dwell position at 10.0 mm, with (a) the new control software and (b) the old control software.

Figure 2

Figure 3. Position and speed of source during transit from first to second dwell positions, from a series of three dwells at 10.0, 15.0 and 20.0 mm, with new control software.

Figure 3

Table 1. Comparison of actual calculated transit dose and EZ BEBIG Multisource® compensated transit dose, for each dwell position in a series of three dwells, at 10.0, 15.0 and 20.0 mm, evaluated at D10 and D20 (10.0 and 20.0 mm from the centre of the dwell position perpendicular to the source movement axis), for the apparent activity of a new Ir-192 source. The data is presented for both the new and old control system software.

Figure 4

Figure 4. Autoradiographs of actual source dwell positions compared to planned positions (vertical lines) as a function of curvature of the transfer tube, for the 1400 universal applicator (140 cm length), both for the previous EZ BEBIG Multisource® control system software, ‘old’, and the latest software, ‘new’. The autoradiographs are for a straight transfer tube and for bends induced by displacements of the distal end by 10, 30 and 90 cm toward the treatment unit.