A planar microwave sensor for noninvasive detection of glucose concentration using regression analysis

Published online by Cambridge University Press:  05 May 2023

Tilakdhari Singh*
Affiliation:
Department of Electronics & Communication Engineering, Motilal Nehru National Institute of Technology Allahabad, Prayagraj, India
Piyush Kumar Mishra
Affiliation:
Department of Electronics & Communication Engineering, Motilal Nehru National Institute of Technology Allahabad, Prayagraj, India
Aditya Pal
Affiliation:
Department of Electronics & Communication Engineering, Motilal Nehru National Institute of Technology Allahabad, Prayagraj, India
Vijay Shanker Tripathi
Affiliation:
Department of Electronics & Communication Engineering, Motilal Nehru National Institute of Technology Allahabad, Prayagraj, India
*
Corresponding author: Tilakdhari Singh; E-mail: tilakdhari.2020rel08@mnnit.ac.in
Rights & Permissions [Opens in a new window]

Abstract

This paper presents a planar microwave sensor for the noninvasive detection of glucose concentration in diabetic patients. The designed sensor operates from the 3.8 to 6.2 GHz frequency band, which covers the 5.8 GHz Industrial Scientific and Medical (ISM) band. The designed sensor shows a percentage bandwidth of 23.8% with a reflection coefficient (S11) of −50 dB at the resonance frequency of 5.7 GHz. The detection was carried out by varying the relative permittivity of the blood in accordance with the glucose concentration based on the Cole–Cole model. The measured result is calculated in terms of varying resonance frequency with variation in the reflection coefficient |S11| of the designed sensor. The observed frequency shift and corresponding sensitivity of the sensor are found at 1.7 GHz and 0.089 MHz/mg dL−1, respectively. An experimental validation has also been performed, and the frequency shift is analyzed by interacting the human thumb with the sensor. The simulated and experimental results of the designed sensor suggest that it can be useful for detecting glucose concentration noninvasively for diabetic patients.

Information

Type
RFID and Sensors
Copyright
© The Author(s), 2023. Published by Cambridge University Press in association with the European Microwave Association.
Figure 0

Table 1. The comparison of noninvasive glucose measuring devices using RF sensors

Figure 1

Figure 1. Measurement of dielectric constant at different glucose concentrations in the frequency range of 1–20 GHz.

Figure 2

Figure 2. (a) Schematic and (b) simulated view of the noninvasive model of RF sensor for glucose measurement.

Figure 3

Table 2. Electrical characteristics of the phantom at 5.7 GHz [28]

Figure 4

Figure 3. Dimension of the proposed sensor for noninvasive glucose measurement: (a) patch and (b) ground.

Figure 5

Figure 4. Fabricated antenna view of the back and front sides: (a) ground and (b) patch.

Figure 6

Table 3. Optimized dimensions of proposed planner antenna–based sensor

Figure 7

Figure 5. Simulated results. (a) Reflection coefficient |S11| plot with respect to frequency and (b) comparison of reflection coefficient |S11| of the sensor with and without tissue layer.

Figure 8

Figure 6. Parametric analysis of phantom (a) for phantom position from the port and (b) phantom height from the sensor surface.

Figure 9

Figure 7. Simulated results: (a) variation in reflection coefficient S11 for relative permittivity and (b) close view of frequency shift as per variation in relative permittivity.

Figure 10

Figure 8. (a) Variation in S11 versus conductivity. (b) Variation in S11 versus loss tangent.

Figure 11

Figure 9. Experimental setup for sensor measurement: (a) sensor in free space, (b) sensor with the thumb (human tissues), and (c) sensor with glucose solution connected to VNA.

Figure 12

Figure 10. Comparison of simulated and fabricated proposed antenna results with and without loaded tissues.

Figure 13

Figure 11. Graph of frequency variation at a different level of glucose concentration in (mg/dL).

Figure 14

Figure 12. Regression analysis: graph of the fitted curve for linearity in frequency.

Figure 15

Figure 13. Variation in blood glucose concentration and respective resonance frequency with respect to time.

Figure 16

Table 4. Comparison of the proposed sensor with another existing sensor for blood glucose sensing applications