Use of Mullins sheath to prevent prolapse of Gore Cardioform ASD Occluder in a child with severe scoliosis

Published online by Cambridge University Press:  02 December 2020

Christopher Herron
Affiliation:
Division of Cardiology, Children’s Hospital of Michigan, Detroit, MI, USA
Daisuke Kobayashi*
Affiliation:
Department of Pediatrics, Central Michigan University College of Medicine, Mount Pleasant, MI, USA
*
Author for correspondence: Daisuke Kobayashi, MD, MPH, Division of Cardiology, Children’s Hospital of Michigan, 3901 Beaubien Blvd, Detroit, MI 48201-2119, USA; Department of Pediatrics, Central Michigan University College of Medicine, Mount Pleasant, MI, USA. Tel: +1 (313) 745 5481; Fax: +1 (313) 993 0894. E-mail: dkobayas@dmc.org
Rights & Permissions [Opens in a new window]

Abstract

Transcatheter device closure of an atrial septal defect (ASD) may require an additional supportive technique in challenging cases. We report a 15-year-old male with moderate-sized ASD and severe scoliosis. In spite of adequate positioning of the Gore Cardioform ASD occluder, the device was pulled into the right atrium by a retrieval cord due to the acute release of strong tension between the delivery catheter and its device upon locking. This phenomenon was prevented by the use of Mullins sheath, resulting in a successful release of the device. The use of a Mullins sheath may be considered to deliver a Gore Cardioform ASD device in selected cases.

Information

Type
Brief Report
Copyright
© The Author(s), 2020. Published by Cambridge University Press
Figure 0

Figure 1. Fluoroscopy imaging at left anterior oblique 60° view. (a) A 32 mm Gore Cardioform ASD Occluder is delivered at the atrial septum. There are adequate splays of the device leaflets (arrows). (b) After the release of locking loop, the device is pulled into the right atrium by the retrieval cord. (c) A 37 mm Gore Cardioform ASD Occluder is delivered at the atrial septum. The delivery catheter is through the 12 Fr Mullins sheath (dotted arrow). (d) After the release of locking loop, the delivery catheter remains in the same position due to the support from the Mullins sheath.

Figure 1

Figure 2. Comparison of angles of the Gore Cardioform ASD Occluder and its delivery catheter with and without a Mullins sheath.

Herron and Kobayashi supplementary material

Herron and Kobayashi supplementary material 1

Download Herron and Kobayashi supplementary material(Video)
Video 3.8 MB
Supplementary material: File

Herron and Kobayashi supplementary material

Herron and Kobayashi supplementary material 2

Download Herron and Kobayashi supplementary material(File)
File 12.2 KB

Herron and Kobayashi supplementary material

Herron and Kobayashi supplementary material 3

Download Herron and Kobayashi supplementary material(Video)
Video 15.1 MB