Development of a 70 MHz unit for hyperthermia treatment of deep-seated breast tumors

Published online by Cambridge University Press:  24 May 2017

Johannes Crezee*
Affiliation:
Department of Radiation Oncology, Academic Medical Center, 1105AZ Amsterdam, The Netherlands. Phone: +31 20 5664231
Geertjan Van Tienhoven
Affiliation:
Department of Radiation Oncology, Academic Medical Center, 1105AZ Amsterdam, The Netherlands. Phone: +31 20 5664231
Merel W. Kolff
Affiliation:
Department of Radiation Oncology, Academic Medical Center, 1105AZ Amsterdam, The Netherlands. Phone: +31 20 5664231
Jan Sijbrands
Affiliation:
Department of Radiation Oncology, Academic Medical Center, 1105AZ Amsterdam, The Netherlands. Phone: +31 20 5664231
Gerard Van Stam
Affiliation:
Department of Radiation Oncology, Academic Medical Center, 1105AZ Amsterdam, The Netherlands. Phone: +31 20 5664231
Sabine Oldenborg
Affiliation:
Department of Radiation Oncology, Academic Medical Center, 1105AZ Amsterdam, The Netherlands. Phone: +31 20 5664231
Elisabeth D. Geijsen
Affiliation:
Department of Radiation Oncology, Academic Medical Center, 1105AZ Amsterdam, The Netherlands. Phone: +31 20 5664231
Maarten C.C.M. Hulshof
Affiliation:
Department of Radiation Oncology, Academic Medical Center, 1105AZ Amsterdam, The Netherlands. Phone: +31 20 5664231
Henny P. Kok
Affiliation:
Department of Radiation Oncology, Academic Medical Center, 1105AZ Amsterdam, The Netherlands. Phone: +31 20 5664231
*
Corresponding author: J. Crezee Email: h.crezee@amc.uva.nl
Rights & Permissions [Opens in a new window]

Abstract

A dedicated hyperthermia (HT) system was designed for tumors in intact breast extending beyond the heating depth of our superficial 434 MHz antennas, consisting of a treatment bed fitted with a 50 cm × 40 cm × 16 cm temperature controlled open water bolus. The patient lies in prone position with the breast immersed in the water positioned in front of a 34 cm × 20 cm 70 MHz waveguide operating in the TE10 mode. E-field patterns were measured in a tissue-mimicking phantom. HT was applied once a week with the 70 MHz applicator for six patients treated with thermoradiotherapy for deep lesions of recurrent breast cancer or melanoma. Two 14-sensor thermocouple thermometry probes were placed in catheters to monitor the invasive temperature. Results: Phantom measurements showed sufficient penetration depth up to 10 cm depth. The combination of 300–900 W antenna power and a water temperature of 42°C was well tolerated for the entire session of 1 h and resulted in good tumor temperatures with T90 = 39.8°C, T50 = 41.1°C, and T10 = 42.2°C. No toxicity or complaints were associated with the heating. A water mattress and other measures were needed to assure a comfortable position throughout the treatment. Conclusion: the 70 MHz breast applicator system performed well and tumor temperatures were good.

Information

Type
Research Papers
Copyright
Copyright © Cambridge University Press and the European Microwave Association 2017 
Figure 0

Fig. 1. Treatment setup: patient in prone position, breast immersed in open water bolus. 70 MHz waveguide placed at the bottom.

Figure 1

Fig. 2. Setup to determine E-field patterns of the 70 MHz waveguide in the transversal (XY) and sagittal (ZY) midplanes in the liquid saline phantom.

Figure 2

Table 1. Patient characteristics and treatment parameters.

Figure 3

Fig. 3. Waveguide faces breast with tumor, healthy breast is isolated.

Figure 4

Fig. 4. E-field patterns measured in the XY midplane with the setup of Fig. 2.

Figure 5

Fig. 5. Invasive lateral (l) and cranial–caudal (c–c) thermocouple probes placed in catheters in the breast of patient 1.

Figure 6

Fig. 6. CT showing the target region and the location of the invasive lateral (l) and cranial–caudal (c–c) thermocouple probes of Fig. 5, as well as the superficial thermocouple probes on the skin surface (skin).

Figure 7

Fig. 7. Steady-state temperature profile in the lateral catheter of patient 1.

Figure 8

Fig. 8. Steady-state temperature profile in the cranial–caudal catheter of patient 1.

Figure 9

Fig. 9. Temperature achieved averaged over the four sessions per patient.

Figure 10

Fig. 10. Power during treatment averaged over the four sessions per patient.