Transcatheter retrieval of a fully dislodged atrial septal defect occluder from the descending aorta using the “retained super-stiff guide wire” technique

Published online by Cambridge University Press:  15 July 2026

Jiawang Xiao
Affiliation:
Department of Congenital Heart Disease, General Hospital of Northern Theatre Command, Shenyang, China
Jingsong Geng
Affiliation:
Department of Congenital Heart Disease, General Hospital of Northern Theatre Command, Shenyang, China
Jianmin Wang
Affiliation:
Department of Congenital Heart Disease, General Hospital of Northern Theatre Command, Shenyang, China
Zhongchao Wang
Affiliation:
Department of Congenital Heart Disease, General Hospital of Northern Theatre Command, Shenyang, China
Qiguang Wang*
Affiliation:
Department of Congenital Heart Disease, General Hospital of Northern Theatre Command, Shenyang, China
*
Corresponding author: Qiguang Wang; Email: wqg1993@126.com
Rights & Permissions [Opens in a new window]

Abstract

Device embolisation is one of the most feared early complications of percutaneous atrial septal defect closure. We report a 40-year-old woman in whom a 12-mm atrial septal defect occluder was accidentally displaced into the descending aorta 24 hours after an apparently successful procedure. The dislodged device was retrieved percutaneously through a 12-F long sheath that was continuously railroaded over a super-stiff guide wire. This “retained super-stiff guide wire” technique secured sheath stability, avoided aortic wall trauma, and allowed the occluder to be slipped into the sheath without change of its orientation. A larger 16-mm small-waist–big-edge device was subsequently implanted uneventfully. The technical details and advantages of keeping the sheath constantly over a super-stiff guide wire are highlighted to facilitate safe trans-arterial retrieval of embolised cardiac devices.

Information

Type
Case Report
Copyright
© The Author(s), 2026. Published by Cambridge University Press
Figure 0

Figure 1. TTE before and after transcatheter ASD closure. (a–c) Pre-procedure: (a) parasternal short-axis view of the great arteries, (b) apical four-chamber view, and (c) subcostal bicaval view demonstrating the ASD and left-to-right shunt. (d–f) After the first device deployment: (d) parasternal short-axis, (e) apical four-chamber, and (f) subcostal bicaval views documenting the initial device position and no residual shunt. (h–j) After retrieval and implantation of a larger device: (h) parasternal short-axis, (i) apical four-chamber, and (j) subcostal bicaval views confirming stable device position with no residual shunt. White arrow, atrial septal defect; red arrow, atrial septal defect occluder.

Figure 1

Figure 2. Intraprocedural fluoroscopy during transcatheter ASD closure. (a) Postero-anterior (PA) and (b) left anterior oblique (LAO) views showing satisfactory position of the initial 12-mm device. (c) Device embolisation into the descending aorta. (d–i) Step-wise fluoroscopic frames demonstrating percutaneous snare retrieval of the embolised occluder via the femoral artery. (j) Immediate image after delivery of the 16-mm replacement device. (k) PA and (l) LAO views confirming stable final position of the 16-mm “small-waist–large-disc” device without residual shunt. Red arrow, atrial septal defect occluder.