A Continuous-Flow Gas Interface of a Thermal/Optical Analyzer With 14C AMS for Source Apportionment of Atmospheric Aerosols

Published online by Cambridge University Press:  07 November 2016

Konstantinos Agrios
Affiliation:
Department of Chemistry and Biochemistry, University of Bern, Bern, Switzerland Oeschger Centre for Climate Change Research, University of Bern, Bern, Switzerland Laboratory of Radiochemistry, Paul Scherrer Institute (PSI), Villigen, Switzerland
Gary Salazar*
Affiliation:
Department of Chemistry and Biochemistry, University of Bern, Bern, Switzerland Oeschger Centre for Climate Change Research, University of Bern, Bern, Switzerland
Sönke Szidat
Affiliation:
Department of Chemistry and Biochemistry, University of Bern, Bern, Switzerland Oeschger Centre for Climate Change Research, University of Bern, Bern, Switzerland
*
*Corresponding author. Email: gary.salazar@dcb.unibe.ch.
Rights & Permissions [Opens in a new window]

Abstract

This article reports on the performance of a continuous-flow interface for CO2 gas feeding into the ion source of a 200-kV accelerator mass spectrometer (AMS) in splitless mode. Distinct CO2 fractions and subfractions produced by a commercial Sunset OC/EC (organic carbon/elemental carbon) analyzer from ambient atmospheric aerosols filters are injected into the source and then analyzed by their 14C/12C ratio in real time. New features are revealed from organic aerosol subfractions, which thermally desorb close to each other and are only visible by real-time analysis. An optimized setup for this purpose is presented for the measurement of CO2 amounts from 3 to 45 μg C. Efficiencies of 4.5–8.0% for the formation of C- ions from CO2 are obtained for sample masses of 5–10 μg C and carbon mass flow rates below 2 µg min–1. However, this ionization efficiency is substantially suppressed at high carbon mass flow rates. The potential of this method for a refined apportionment of aerosol sources is demonstrated with ambient filter samples.

Information

Type
Advances in Physical Measurement Techniques
Copyright
© 2016 by the Arizona Board of Regents on behalf of the University of Arizona 
Figure 0

Table 1 14C/12C analysis of IAEA standards (Rozanski et al. 1992; Le Clercq et al. 1998) using the continuous-flow method with 1σ uncertainties for n=3 repetitions for each material. Measured carbon amounts ranged between 15 and 20 μg C. The P value describes the level of significance from a t test comparing nominal and measured F14C, indicating a statistically significant difference for P values<0.05.

Figure 1

Figure 1 12C– currents from the analysis of HOxII: (a) 6 μg C transported to the ion source by using a 0.13-mm-ID inlet capillary at 2.5 mL min–1 (dashed line) and 3.5 mL min–1 (solid line) He carrier flow rate; and (b) 6 μg C (dashed black line), 10 μg C (red solid line), and 20 μg C delivered (dotted line) via a 0.10-mm-ID inlet capillary and at a constant He carrier flow rate of 1.5 mLmin–1. Corresponding average carbon mass flow rates are indicated in the legend.

Figure 2

Figure 2 Carbon mass flow rate of HOxII depending on (a) the C mass injected at 1.5 mL min–1 and (b) the inverse of the He flow using 6 μg C and He flows of 0.5–3.5 mL/min. Measurements with ionization suppression are indicated by crosses. In all measurements, CO2 was transported to the ion source using a 0.10-mm-ID inlet capillary. Same experimental details as in Figure 1.

Figure 3

Figure 3 Influence of the carbon mass flow rate on (a) ionization yields using fixed sample masses of 6 µg C and variable He flows and (b) 12C– currents depending on the mass of the sample and geometry of the gas target. Measurements with ionization suppression are indicated by crosses. Same experimental details as in Figure 1.

Figure 4

Figure 4 Continuous-flow 14C/12C measurement overlaid with the 12C– current and the 13C/12C ratio for an aerosol filter from Zurich from January 2008 treated with the OC/EC protocol consisting of three temperature ramps (375, 475, and 650°C) under pure oxygen. The elevated baseline after the second peak is due to an artifact of the measurement and has not been reproduced in other runs.

Figure 5

Figure 5 Continuous-flow 14C/12C analysis of OC subfractions (a) evolving at 200, 260, 300, and 375°C for an aerosol filter from Zurich for 28/01/2009; and (b) of OC subfractions evolving at 200 and 375°C for an aerosol filter from the Los Angeles Basin for 20/05/2010. Please note that both examples are also included in Table 2.

Figure 6

Table 2 14C real-time analysis of a selection of ambient aerosol filters from several campaigns. The F14C values are averaged over distinct CO2 subfractions evolving at 200, 260, 300, and 375°C or 200 and 375°C. Each filter was measured only once. The uncertainty is related to the counting statistics. Statistically significant 14C shifts within such subfractions are indicated with an asterisk. For the filters from the Los Angeles Basin, three 1.5-cm2 quartz filter punches were piled up in the sample holder during combustion. ND means not detected.