Hostname: page-component-74656f9884-j7cmm Total loading time: 0 Render date: 2026-11-08T14:15:47.004Z Has data issue: false hasContentIssue false

Navigating the challenges of misplaced electrode array in cochlear implant surgery: causes, consequences and solutions

Published online by Cambridge University Press:  02 April 2025

Mohnish Grover
Affiliation:
Department of Otorhinolaryngology, SMS Medical College, Jaipur, RJ, India
Tanmaya Kataria*
Affiliation:
Department of Otorhinolaryngology, Mahatma Gandhi Medical College, Jaipur, RJ, India
Sunil Samdani
Affiliation:
Department of Otorhinolaryngology, SMS Medical College, Jaipur, RJ, India
Gaurav Gupta
Affiliation:
Department of Otorhinolaryngology, SP Medical College, Bikaner, RJ, India
Raghav Mehta
Affiliation:
Department of Otorhinolaryngology, RUHS College of Medical Sciences, Jaipur, RJ, India
Mukesh Dagur
Affiliation:
Department of Otorhinolaryngology, RDBP Jaipuria Hospital, Jaipur, RJ, India
*
Corresponding author: Tanmaya Kataria; Email: tanmayakataria@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

Objectives

Cochlear implantation is a groundbreaking surgical procedure offering significant benefits to individuals with severe hearing impairments. Correct placement of the electrode array within the cochlea is crucial for optimal device functionality and auditory outcomes. Our study describes four cases of electrode misplacement along with the factors causing it and discusses solutions for the same.

Methods

Eight hundred cochlear implant cases were reviewed in three tertiary care centres in India, spanning 2011 to 2023.

Results

Four cases of electrode misplacement requiring reimplantation were identified.

Conclusion

This paper discusses the implications of electrode misplacement, the factors contributing to it and the critical role of multidisciplinary approaches in managing such complications. By addressing these issues, we aim to improve surgical techniques and outcomes in cochlear implant procedures.

Information

Type
Main Article
Copyright
© The Author(s), 2025. Published by Cambridge University Press on behalf of J.L.O. (1984) LIMITED.
Figure 0

Figure 1. (a) Transorbital viewradiograph showing lack of standard spiral configuration of electrode array (red arrow). (b) Axial section of HRCT temporal bone showing presenceof electrode array in the middle ear (red arrow).

Figure 1

Figure 2. (a) Lack of standard spiral configuration of electrode array (yellow arrow) on modified Stenvers view radiograph of the temporal bone is suggestive of electrode misplacement. (b) Axial sections of HRCT right temporal bone show part of electrode array indicated by black arrow in the middle ear (A) that passed tangentially from cochlear basal turn (B) and then entered into the internal carotid artery canal as shown by red arrow (C, D and E).

Figure 2

Figure 3. Coronal sections of HRCT left temporal bone show part of electrode array indicated by red arrow coiled in the hypotympanum.

Figure 3

Figure 4. Coronal sections of HRCT right temporal bone show electrode array traversing through the tunnel made in external auditory canal and then deflected into the superior semicircular canal indicated by red arrow.

Figure 4

Figure 5. Sketch diagram of a posterior tympanotomy showing electrode insertion through round window. Magnified view of round window shows A) deflection of electrode array towards scala media/vestibuli after hitting crista fenestrae then B) after drilling crista fenestrae straight insertion of electrode array in anteroinferior direction in scala tympani can be seen.

Figure 5

Figure 6. Axial section of HRCT right temporal bone shows infracochlear air cell track indicated by the red arrow which can be mistakenly identified as the round window and smooth insertion of electrode can mislead the surgeon.