Hostname: page-component-7b7567c8c-7pwfn Total loading time: 0 Render date: 2026-11-15T11:02:30.851Z Has data issue: false hasContentIssue false

Size effects in indentation of hydrated biological tissues

Published online by Cambridge University Press:  26 October 2011

Michelle L. Oyen*
Affiliation:
Department of Cambridge University Engineering, Cambridge CB2 1PZ, United Kingdom
Tamaryn A.V. Shean
Affiliation:
Department of Cambridge University Engineering, Cambridge CB2 1PZ, United Kingdom
Daniel G.T. Strange
Affiliation:
Department of Cambridge University Engineering, Cambridge CB2 1PZ, United Kingdom
Matteo Galli
Affiliation:
Department of Cambridge University Engineering, Cambridge CB2 1PZ, United Kingdom; and Laboratoire de Mécanique Appliqée et d’Analyse de Fiabilité, École Polytechnique Fédérale Lausanne, Lausanne 1015, Switzerland
*
a)Address all correspondence to this author. e-mail: mlo29@cam.ac.uk

Abstract

Fluid flow in biological tissues is important in both mechanical and biological contexts. Given the hierarchical nature of tissues, there are varying length scales at which time-dependent mechanical behavior due to fluid flow may be exhibited. Here, spherical nanoindentation and microindentation testings are used for the characterization of length scale effects in the mechanical response of hydrated tissues. Although elastic properties were consistent across length scales, there was a substantial difference between the time-dependent mechanical responses for large and small contact radii in the same tissue specimens. This difference was far more obvious when poroelastic analysis was used instead of viscoelastic analysis. Overall, indentation testing is a fast and robust technique for characterizing the hierarchical structure of biological materials from nanometer to micrometer length scales and is capable of making quantitative material property measurements to do with fluid flow.

Information

Type
Articles
Copyright
Copyright © Materials Research Society 2011
Figure 0

TABLE I.List of indentation experiments. For each set of experiments, the table gives the specimen type, the indentation testing machine (see text for descriptions), the spherical probe radius, the peak load or peak displacement level, load control (LC) or displacement control (DC), the ramping and holding times for the experiment, and the total number of individual indentation tests for which data are presented here.

Figure 1

FIG. 1.Instantaneous shear modulus (G0) and viscoelastic ratio (G∞/G0) for a PS-4 polymer tested by nanoindentation (nI) and microindentation (μI) as a function of tip radius for seven different indenter tips in displacement control (solid symbols). Viscoelastic properties of the PS-4 polymer obtained on a different machine in load control with a sapphire tip (using only data in a similar load range) are included for reference (open symbols). There is no effect of tip radius, machine type, or control mode on either the elastic or viscoelastic response for the PS-4 polymer.

Figure 2

FIG. 2.Instantaneous shear modulus (G0) and viscoelastic ratio (G∞/G0) for cartilage specimens tested by nanoindentation (nI) or microindentation (μI), plotted as a function of mean indentation strain. Differences are observed for both elastic and time-dependent responses as a function of length scale and depending on the experimental control mode used.

Figure 3

FIG. 3.Instantaneous shear modulus (G0) and viscoelastic ratio (G∞/G0) for bone specimens tested by nanoindentation (nI) or microindentation (μI). The elastic moduli did not vary, but there was a slight difference in viscoelastic ratio.

Figure 4

FIG. 4.Poroelastic parameters identified from spherical nanoindentation and microindentation tests on all bone specimens; the value of the undrained Poisson?s ratio (νu) was fixed at 0.5. The elastic properties (G, ν) were invariant but a substantial difference between permeability values was observed when comparing nanoindentation (nI) and microindentation (μI) results.

Figure 5

FIG. 5.Poroelastic parameters identified from spherical nanoindentation and microindentation tests on a single bone specimen. The undrained Poisson's ratio (νu) was assigned a value of 0.4, 0.45 or 0.5 to assess sensitivity of the parameter identification process on this value.

Figure 6

FIG. 6.Optical micrograph of an indented bone surface, showing the characteristic size of the indentation contacts for microindentation (dashed lines, for which the inner circumference is for an applied load of 25 N and the outer for a load of 50 N) and nanoindentation (filled circle).

Figure 7

TABLE II.Bone nanoindentation results comparison.

Figure 8

FIG. 7.Poroelastic parameters for spherical nanoindentation of cartilage specimens from the current study, compared to nanoindentation and confined and unconfined compression performed by Miller and Morgan,17 and flat punch microindentation performed by Mow39 (1989). The undrained Poisson’s ratio (νu) was fixed at 0.5. The elastic properties (G, ν) were invariant, but a large difference between permeability (k) values was observed when comparing both sets of nanoindentation (nI) and microscale (μC, μI) results. *Plotted standard deviation bars were based on 5% significance levels, as used in the article.