IMPROVING IONPLUS MICADAS PERFORMANCE WITH RECESSED GRAPHITE

Published online by Cambridge University Press:  04 April 2024

Taylor A B Broek*
Affiliation:
National Ocean Sciences Accelerator Mass Spectrometry Facility, Woods Hole Oceanographic Institution, 266 Woods Hole Rd., Woods Hole, MA, USA
Mark L Roberts
Affiliation:
National Ocean Sciences Accelerator Mass Spectrometry Facility, Woods Hole Oceanographic Institution, 266 Woods Hole Rd., Woods Hole, MA, USA
*
*Corresponding author. Email: taylor.broek@whoi.edu
Rights & Permissions [Opens in a new window]

Abstract

Cathodes with recessed sample surfaces have several benefits in cesium sputter ion sources, including higher output, more efficient use of sample material, and improved focusing of the extracted ion beam. However, the Ionplus MICADAS uses cathodes with a graphite surface that is essentially flush with the sample holder. To evaluate the performance of recessed graphite with the MICADAS and determine the optimal surface depth, we tested four different depths, including the standard (flush) pressing method, 0.5 mm, 1.0 mm, and 1.5 mm. We found that recessed depths of 1.0 and 1.5 mm resulted in 20% higher ion beam current compared to the standard method under the same source conditions. The results are consistent with the beam produced from the recessed targets being more narrowly focused with a lower emittance, resulting in better transmission through the accelerator. Small graphite samples (200 µg C) with recessed surfaces produced higher currents for longer, leading to a 2–3× increase in sample ionization efficiency. Additionally, there was some evidence that isotopic ratio measurements of recessed samples were more stable over time. Overall, samples recessed to 1 mm depth offered numerous advantages over the standard pressing method and we have subsequently started pressing all MICADAS graphite using this approach.

Information

Type
Conference Paper
Copyright
© Woods Hole Oceanographic Institution, 2024. Published by Cambridge University Press on behalf of University of Arizona
Figure 0

Figure 1 Target holder assembly of the Ionplus pneumatic sample press shown both without (A) and with (B) target. (C) Custom built inserts for producing targets with graphite surfaces recessed to the depths described in this study. The exposed section of the drill stems are 0.5, 1.0, and 1.5 mm in length. (D) 1 mm depth insert showing set screw for holding drill stem. (E-G) Modified target holder assembly shown with and without pressing pin and target.

Figure 1

Figure 2 Current output (measured as µA of low-energy 12C–) from graphite targets containing 1 mg C pressed to different depths. Each point represents the average current of n=3 separate targets measured over a 180-second pass (15 cycles of 12 seconds each).

Figure 2

Figure 3 Left panel: current output (measured as µA of low-energy 12C–) from graphite targets containing 200 µg C pressed to different depths. Each point represents the average current of a single target measured over a 180-second pass (15 cycles of 12 seconds each). Shaded region represents measurements excluded from the efficiency calculation. Right panel: summed total conversion of C to C– of each target during measurement passes where the average 12C– current was greater than 40 µA. The number in parentheses above each bar denotes the number of passes with an average current above 40 µA that were used in efficiency calculations.

Figure 3

Figure 4 Transmission (measured as the ratio of high-energy 12C+ to low-energy 12C–) from graphite targets containing 200 µg C and 1 mg C pressed to different depths. Each bar represents the average transmission of n=3 1 mg C targets (dashed bars) or a single 200 µg C target (solid bars) measured over thirty 180-second passes (15 cycles of 12 seconds each).