Assessment of normal tissue objectives in RapidArc treatment for cervical cancer

Published online by Cambridge University Press:  09 May 2024

Athulya P.*
Affiliation:
Medical Physics, Kalyan Singh Super Specialty Cancer Institute, Lucknow, Uttar Pradesh, India
Sumanta Manna
Affiliation:
Medical Physics, Kalyan Singh Super Specialty Cancer Institute, Lucknow, Uttar Pradesh, India
Sharad Singh
Affiliation:
Radiation Oncology, Kalyan Singh Super Specialty Cancer Institute, Lucknow, Uttar Pradesh, India
*
Corresponding author: Athulya P.; Email: athulyamedphy@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

Purpose:

The normal tissue objective (NTO) is a tool used in inverse-planned intensity-modulated radiation therapy (IMRT) to minimize dose dispersion to surrounding tissues. The current study focuses on the NTO’s impact on RapidArc treatment plans for cervical cancer patients or its role in reducing doses to healthy surrounding tissues.

Material and methods:

This study included 11 cervical cancer patients who underwent RapidArc treatment. We assessed plans both with and without the NTO objective by evaluating parameters such as homogeneity, conformity, gradient index (GI), IMRT factor, integral dose and the volume of normal tissues receiving low doses of 40, 30, 20 and 10 Gy. Further, differences between automatic NTO and manual NTO were evaluated using Wilcoxon signed-rank test.

Results:

There were no significant differences in the conformity index, homogeneity index, IMRT factor and integral dose between plans with automatic NTO and those with manual NTO RapidArc plans. However, we did observe a clear advantage in using manual NTO for controlling low-dose exposure to normal tissues. The comparisons between automatic and manual NTO resulted in a GI of 3·1 ± 0·3 versus 2·7 ± 0·68 (p = 0·008). Furthermore, we noticed a significant reduction in the volumes receiving low doses (V10, V20, V30 and V40) with the manual NTO settings.

Conclusion:

The NTO plays a crucial role in optimizing RapidArc plans for treating cervical cancer. Based on the findings of this study, manual NTO settings of distance from PTV border xstart = 0·5 mm, start dose f0 = 105%, end dose f∞ = an average of 40%, dose fall-off 0·2 mm–1 were optimal. Further research involving a larger sample size and exploration of various NTO parameters is necessary to validate our results.

Information

Type
Original Article
Creative Commons
Creative Common License - CCCreative Common License - BYCreative Common License - NCCreative Common License - ND
This is an Open Access article, distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivatives licence (http://creativecommons.org/licenses/by-nc-nd/4.0/), which permits non-commercial re-use, distribution, and reproduction in any medium, provided that no alterations are made and the original article is properly cited. The written permission of Cambridge University Press must be obtained prior to any commercial use and/or adaptation of the article.
Copyright
© The Author(s), 2024. Published by Cambridge University Press
Figure 0

Figure 1. Dialog window from Eclipse for normal tissue objective (NTO) setting.

Figure 1

Figure 2. Adjustment criteria for automatic NTO. (1) Dose. (2) Distance from the target. (3) The dose at this point is above the accepted dose level, and therefore, the optimization tries to reduce the dose in this area. (4) Accepted dose level for this case. (5) Points within this region are not affected.

Figure 2

Figure 3. The variation of PTV coverage (V95 and V98 for various priorities using automatic NTO).

Figure 3

Table 1. Evaluation of plan indices (CI, HI, GI and IMRT factor) and integral dose for automatic normal tissue objective (NTO) with different priority.

Figure 4

Figure 4. The variation of low-dose volumes (V40, V30, V20 and V10) of normal tissues for automatic NTO with different priority.

Figure 5

Table 2. Evaluation of plan indices (CI, HI, GI and IMRT factor) and integral dose with NTO for different fall-off setting and without NTO.

Figure 6

Figure 5. The variation of PTV coverage (V95 and V98) for various fall-off values using manual NTO and without NTO.

Figure 7

Figure 6. The variation of low-dose volumes of normal tissues for manual NTO and without NTO.

Figure 8

Table 3. Comparison of manual and automatic NTO for different plan indices (HI, CI, GI and IMRT factor), low-dose volume (V40, V30, V20 and V10) and integral dose.