To save content items to your account,
please confirm that you agree to abide by our usage policies.
If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account.
Find out more about saving content to .
To save content items to your Kindle, first ensure no-reply@cambridge.org
is added to your Approved Personal Document E-mail List under your Personal Document Settings
on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part
of your Kindle email address below.
Find out more about saving to your Kindle.
Note you can select to save to either the @free.kindle.com or @kindle.com variations.
‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi.
‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.
SGRT can accurately monitor VMAT DIBH breast radiotherapy patients. Camera occlusion may provide difficulties for SGRT systems, such as VisionRT’s AlignRT, during rotational imaging (CBCT) and treatments (VMAT). Therefore, appropriate regions of interest (ROIs) need definition to ensure robust monitoring. Equally, operators must know where camera occlusion most likely occurs to optimise workflows.
Methods:
A CT scan of an anthropomorphic phantom was performed to mimic a breast patient’s DIBH treatment position. A VMAT plan was produced, and six ROIs drawn on an AlignRT terminal. An ROI sensitivity test was performed, and approximate percentage surface area of occlusion was determined for all ROIs. Postural video was also utilised to determine where camera occlusion is most likely to occur.
Results:
No full camera occlusion was demonstrated using the phantom. However, postural video demonstrated camera occlusion was more probable between 10 and 60 degrees for CBCT arcs due to lateral camera occlusion. Sensitivity test results showed ROIs were more sensitive to vertical and lateral translations when both lateral cameras are blocked at CBCT50. Larger ROIs may be less sensitive and more robust when cameras were occluded. Additionally, the greatest approximate percentage of occlusion occurred at the superior lateral aspects of ROIs.
Conclusion:
Gantry angles at higher risk of camera occlusion have been described between 10 and 60 degrees for CBCT to assist operators with troubleshooting, and a clinical workflow proposed. An ROI for VMAT DIBH breast rotational imaging and treatment has also been suggested to reduce the impact of camera occlusion.