To save content items to your account,
please confirm that you agree to abide by our usage policies.
If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account.
Find out more about saving content to .
To save content items to your Kindle, first ensure no-reply@cambridge.org
is added to your Approved Personal Document E-mail List under your Personal Document Settings
on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part
of your Kindle email address below.
Find out more about saving to your Kindle.
Note you can select to save to either the @free.kindle.com or @kindle.com variations.
‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi.
‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.
Implantable haemodynamic monitors allow remote monitoring of Fontan circulation. We report unique opportunities and challenges related to device use in rural, high-altitude regions.
Objectives:
Assess the performance of implantable haemodynamic monitor in Fontan circulation and identify potential sources of measurement discrepancy defined as non-physiological, negative, or significantly lower reading than baseline.
Methods:
We performed a retrospective review of patients who underwent implantable haemodynamic monitor implantation from September 2021 to April 2024 (n = 17) at our centre (∼1,000 feet above sea level; ASL) and identified those with sensor discrepancies.
Results:
During a mean follow-up duration of 26 months (range 13–44 months), there were no procedure-related complications, thromboembolism, or device displacement. Ten patients lived in rural, higher-altitude regions (average altitude 5100 feet above sea level, average distance from centre ∼160 miles, range = 100–400 miles). Challenges in remote monitoring included unreliable home-internet connection, non-compliance, and difficulty performing device recalibration at patient’s home altitude. Sensor discrepancies were noted in 7 patients (41%), of whom 6 (86%) lived remotely. Manual review of the waveforms identified sources of discrepancy, including misinterpretation of the non-pulsatile pressure waveform (n = 3), offset due to change in hospital-interrogation unit (n = 4), and sensor drift (n = 1). Altitude change did not directly affect sensor performance. We were able to apply corrective interventions in 4/7 sensors, including Fontan-specific settings (overriding pulsatility), and back-end recalibration, which were effective in improving device accuracy.
Conclusions:
Implantable haemodynamic monitors are a promising tool for monitoring Fontan circulation but may require modified settings and careful attention to potential interpretation errors. Home monitoring remains challenging for rural, high-altitude residents with limited resources.