To save content items to your account,
please confirm that you agree to abide by our usage policies.
If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account.
Find out more about saving content to .
To save content items to your Kindle, first ensure no-reply@cambridge.org
is added to your Approved Personal Document E-mail List under your Personal Document Settings
on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part
of your Kindle email address below.
Find out more about saving to your Kindle.
Note you can select to save to either the @free.kindle.com or @kindle.com variations.
‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi.
‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.
Manual contouring (MC) is time-consuming work in radiotherapy planning for rectal cancer. Artificial intelligence (AI) can reduce the time required for clinical target volume (CTV) and organs-at-risk (OARs) delineation. In this study, we evaluated the quality of auto-segmented CTVs and OARs.
Methods:
Dose-planning data were collected from ten patients who underwent preoperative radiotherapy for locally advanced rectal cancer in 2024. Auto-segmented structures from the AI-Rad and Contour+ software tools were added. Constructed AI-CTVs, based on Contour+ segmentations and AI-OARs, i.e., bladder, femoral heads and bowel bag, by both AI tools, were compared to their MC counterparts by use of quantitative metrics, volumetric/surface Dice similarity coefficients (vDSC/sDSC) and maximum/average Hausdorff distance (HD/aHD). The constructed AI-CTVs and MC counterparts were graded by two radiotherapists with two qualitative methods.
Results:
The median vDSC, sDSC, HD and aHD values of our constructed AI-CTVs compared with the MC-CTVs were 0.86, 0.61, 23.19 and 0.62 mm, respectively. For both AI tools, the agreement in the OAR metrics was overall good but less similar for the bowel bag. The qualitative evaluations of the AI-CTVs, compared to the MC-CTVs, were in clear favour of the MC-CTVs. The cranial-anterior nodal levels were anatomical areas with poorer coverage, where the contouring guidelines differed.
Conclusion:
The quality of our constructed AI-CTVs was inferior to the MC-CTVs. Thus, the auto-segmentation methods need further development on this aspect for use in the clinical setting. In contrast, the agreement of the quantitative metrics for the OARs was overall good, except for the bowel bag.
The authors proposed a novel perimeter-based index (PBI) that was capable of evaluating the accuracy in the appraisal of auto-segmentation software. A quantitative value, that is time saved in editing the auto-segmented contours, was used to compare the effectiveness of two other commonly used indices in this study.
Methods
The relationship between the proposed index and the amount of the contouring time that could be saved was studied. The performances of two other commonly used similarity indices, namely Dice similarity coefficient (DSC) and the modified normalised average Hausdorff distance (MNAHD), were also evaluated. Ten nasopharyngeal cases and ten prostate cases that were previously treated with intensity-modulated radiation therapy technique were recruited as the validation cases in this study. Three observers were invited to contour four structures (bladder, rectum, brain stem and parotid gland) on computed tomography images of the validation cases without any aids. The time taken for contouring was recorded as the manual contouring time. By using an atlas-based auto-segmentation software, three sets of contours were generated for each validation case with different library sizes to produce different degrees of similarity level. The values of the three similarity indices of the auto-segmented contours were calculated. The observers were asked to edit the auto-segmented contours and the editing time was recorded.
The correlation between the editing time and the similarity indices was studied. The amount of time saved was calculated by subtracting the editing time from the manual contouring time. The performances of PBI, DSC and MNAHD were evaluated using Pearson correlation coefficient and receiver operating curve (ROC) analysis.
Results
The PBI showed a positive linear relationship with the amount of contouring time saved. Pearson correlation coefficient ranged from 0·73 to 0·86 for the four structures. The PBI had a stronger correlation than the DSC in bladder and parotid gland, while there was no significant difference between the two indices in rectum and brain stem. The MNAHD had an inferior correlation than the proposed index. For the ROC analysis, the cut-off values for the PBI were 0·549, 0·401 and 0·301 for the three levels of contouring time saved, namely 50, 25 and 0%, respectively. The accuracy of PBI was over 77% and the Youden index was >0·6 for all three levels.
Conclusions
The proposed index showed a stronger relationship to the amount of contouring time saved. It was a simple tool that could be used to evaluate the performance of different segmentation algorithms.